- No zero point error
- Very accurate usually better than +/- 0.2mg/l
- Calibrate to 100% in air
- Self-temperature compensating from 5 to
40 deg C
- Connect with good quality two core cable, tested up to
2000m with ordinary cable
- Very stable, usually 3 to 6 months between calibration
checks
- Very easy to maintain, can be serviced in the field
- Strong Teflon membrane
- Very heavy duty sub-sea urethane cable, into epoxy sealed
internals
- Diameter = 63 mm, length = 73 mm. Cable length = 5 m
(standard)
- Weight 1000 g incl. cable
5m
- 6
to 12 mill volt per ppm (mg/l) ( depends on temp)
- pressure to 10 atmospheres
- Water flow should be at least 4 cm/sec
- Response time, to change in DO, approx 15 to 30 seconds
- Standard probe is a 2 wire, with temp sensor there are
5 wires
- Temperature sensor PT100 built in to probe
is available as an option
|
 |
|
Calibration of the oxygen probe is very easy, just follow the
steps below;
Standard calibration
Remove probe from water, dry and clean the probe and
membrane with a soft clean cloth.
Using a pin, clear the breather hole on top of the probe
Suspended the probe in air above the surface of the water,
try to keep the probe out of the wind and direct sun-light. The probe should
be left for a period of at least 30 minutes, or until you are sure that
the temperature of the probe is the same as the air temperature. The
best time to perform this task is early in the morning or late evening.
The probes will now be reading the equivalent of 100%
saturation. Take an accurate reading of the air temperature by the probes. Use
the table below to establish the concentration in mg/l for the probe, and use
this figure with your oxygen monitoring equipment.
Example. if the air temp is 20.5 deg C,
and the probe is at the same temperature as the air, the probe should be
reading 8.99mg/l. Set the display to give this reading
Quick calibration
Sometime you do not have time to wait 30
minutes or more to perform a calibration, we therefore have a quick
calibration technique that often proves more accurate than the standard
technique.
Remove the probe (or probes from
the water), do not dry the probes, suspended in air out of direct sunlight
Clear the breather hole with a pin
and wipe the membrane with a soft cloth
Measure the temperature of the
water ( if more than one probe is being calibrated, the water
temperature should be know for each probe)
Calibrate the probes using the
water temperature. The probes must be calibrated within 5 minutes of
removing from the water, otherwise the temperature of the probe may have
changed too much from the temperature of the water
|
|
|
The oxygen probe requires very little attention, the degree of
attention will depend upon the water type in which the probe is immersed. For most
applications we recommend that the probe is removed from the water once a week and the
membrane cleaned with a soft clean cloth. The breather hole on top of the probe should
also be cleaned using a pin. This is all that you need do with the probe on a regular
basis.
In the event that the membrane is damaged, the readings will become
very erratic. Under these conditions the membrane should be replaced.
Membrane Replacement
The following procedure describes membrane replacement.
Remove the oxygen probe from the water and clean with a cloth or paper
towel
Unscrew the bottom end cap, please note that the electrolyte containing
a white deposit of colloidal solution of zinc oxide may be captained in the
cap. Discard this solution.
Using a coin, unscrew the threaded retaining ring in the membrane cap,
remove and discard the membrane and small `o` ring located below the membrane
Clean the inside of the membrane cap thoroughly with a
damp cloth, and
finish off with a clean dry cloth.
Insert a new `o` ring, and then the membrane on top of the `o`
ring. Screw down the threaded retaining ring until you feel tension, then give is a
further 1/4 turn. If the membrane wrinkles, you will need to replace the `o`
ring and membrane and try again.
Clean the inside of the probe top, you may clean the silver cathode with
1500 grade paper, take care as damage to the silver cathode can affect probe
readings. You can clean the angular zinc anode with rough abrasive paper in order to
remove any oxidation layer. Clean the probe in freshwater to remove any of the fines.
Fill the membrane cap with electrolyte, and holding the probe
vertically, screw the membrane cap on to the top of the probe, making sure that the large
`o` ring is in place. Slowly screw up the cap, the excess electrolyte will escape
though the breather hole. Take care that you do not screw the cap on too quickly
since this will stretch the membrane. Screw up the cap until it seals on the large `o`
ring a then give it a further 1/4 turn.
Immerse the probe into the water, after 24 hours perform a calibration
as per the instructions above.
Components supplied with the probe
The standard probe is supplied with he following components;
Oxygen probe fitted with 5 meters of heavy duty cable
50ml bottle of electrolyte
5 spare membranes and membrane o rings
Note.
The probe is supplied dry without any electrolyte. On receipt of the
probe, remove the end cap, fill end cap with electrolyte and slowly screw the cap back up
onto the probe. Screw the end cap up slowly to allow the excess electrolyte to
squirt out of the top breather hole. You should then leave the probe for 30 minutes
suspended in air before you perform a calibration. Once the probe is calibrated immerse
the probe into the water for a period of at least 24 hours and then perform another
calibration. Thereafter a calibration will be required at approximately 1 to 2 week
intervals. After a period of approx. 4 weeks, the calibration frequency can be
reduced to once every month to several months.
Low Temperature applications
If the oxygen probe is subjected to sub-zero temperatures, the
electrolyte may freeze and cause damage to the probe. In order to avoid freezing
down to temperatures of -25 deg C glycerine is added to the electrolyte at a level of 55%
glycerine to 45% water by volume. The glycerine will reduce the milli volt out-put of the
probe down to approximately 30mv.
|
OxyGen Probe
wire colours
|
|
Integration with your own system
The probe can be considered a low impedance millivolt generator. It has built in
temperature compensation for readings only in mg/l. The output is linearly proportional to
the oxygen concentration, and it is about 120 mv in air at
a temperature of 20 deg C. The out-put is approx 10mv per mg/l of oxygen.
The output impedance is very low (< 1 kohm) so noise problems normally do not exist.
However it is important that the input impedance of the transmitter is at least 3 milli
ohms in order not to disturb the temperature compensation. It is also extremely important
that inputs from probes are galvanically isolated from each other if more than one probe
is connected to the same electronics.
The oxygen probe is easy to install. It should be placed where there is some
movement in the water 2 cm/sec. Ensure that the probe cannot strike against the tank wall
or similar) and don't mount it directly above air or oxygen diffusers etc.
Cable connections
Each probe is connected using ordinary 2-core cable and it is fitted with 5 meters of
urethane cable unless otherwise specified. The brown ( or red) wire is positive the
blue (or white) negative Use the junction box supplied with the probe when extending
the cable.
With regards to the temperature sensor ( type pt100) the cable is numbered as follows;
Standard Oxygen sensor
+ve oxygen wire
(brown)
-ve oxygen wire
(blue)
Oxygen probe with pt100 sensor
1 +ve dissolved
oxygen ( red)
2 -ve dissolved
oxygen connection ( blue)
3 connected to the
other side of the PT100 ( black)
4 & 5 joined together on one side of the PT100 ( green
& yellow)
Suppression Circuit
The suppression circuit is used to drop the
out-put of our oxygen probes from approximately 80mv to around 25mv. The
suppression circuit is required for oxygen meters such as Eutech units.

|
|
Problem
solving |
|
In the unlikely event that you experience a
problem with our oxygen probes, then there are some simple checks that you
can perform to identify and or solve the problem.
|
Problem |
Description |
Action |
|
Zero or low reading |
A zero reading means that there is no
voltage being generated by the probe, or there is a break in the electrical
circuit. The milli volt out-put from the probe should be between 60 and
120mv in most cases, with the probe in air at 20 deg C The out-put is
approx 10mv per mg of oxygen. If the reading is very low check the following.. |
|
| Check that there is electrolyte in the
probe |
change the membrane and electrolyte |
| Check that there is no cable damage |
If there is cable damage then cut out or
repair the damage. Check that there is no water or moisture in junction
boxes or cable connections. |
| biofouling of membrane can occur in in
biologically active water. |
clean the membrane with a soft tissue
once a week, or more frequently if required |
|
High reading |
The milli volt out-put from the probe should be between 60 and
120mv in most cases, with the probe in air at 20 deg C The out-put is
approx 10mv per mg of oxygen. If the reading is higher, check the
following; There should only be one small `o` ring
below the membrane, if there are two `o` rings, or if the `o` ring is above
the membrane then this will give the wrong tension on the membrane.
|
change the membrane and electrolyte |
| The large `o` ring between the top and
bottom sections of the the probe must be fitted, if it is missing then
there will be too high a tension on the membrane. |
change the membrane and electrolyte and
fit a replacement large `o` ring |
| Check the silver cathode for any damage.
the cathode should be clean and smooth without any indentations.
If the cathode has been damage, you will experience high erratic readings |
If there is only minor damage then using
wet/dry paper, 700 grade gently remove the indentations. |
| If none of the above apply, remove the
electrolyte from the probe, wash it electrode in fresh water an then
thoroughly dry. The probe should now give a zero reading which
will be less than 2 mv or 0.2mg/l of oxygen. If the probe gives a
reading above 2 or 3mv, then it means that there is a short out
in the probe due to water ingress. |
If there is no damage to the probe cable,
then the failure is internal. Return the probe to Dryden Aqua. |
| |
|
|
un-stable reading |
New probes |
It takes approximately 6 weeks for a new
probe to fully condition stabilise. |
| Blocked breather hole. All probes have a
small breather hole which should be kept clean. If the hole becomes blocked,
gas inside the probe can inflate the membrane and push it off the cathode.
This can give high and or unstable readings |
clean the hole with a pin once a week. |
| If the prove is hitting the side of a
tank, the readings can become erratic |
position the probe such that it does not
swing and hit the side of the tank. |
| Air or oxygen bubbles sitting below the
membrane |
When the oxygen probe is used in aeration
or oxygenation tanks, bubbles can sit on the underside of the membrane,
this can lead to false readings. Position the probe in a strong current of
water, or tilt the probe by 45 deg. |
|
|
Solubility of Oxygen in water
The solubility of oxygen in water in equilibrium with air at 760mm Hg pressure and 100%
relative humidity Units:mg/1 . The data gives the mg/l of
oxygen in solution that is equivalent to 100% saturation at the specified
temperature
|
| T deg C
|
0.0 |
0.1 |
0.2 |
0.3 |
0.4 |
0.5 |
0.6 |
0.7 |
0.8 |
0.9 |
| 0 |
14.60 |
14.65 |
14.52 |
14.48 |
14.44 |
14.40 |
14.36 |
14.33 |
14.29 |
14.25 |
| 1 |
14.21 |
14.17 |
14.13 |
14.09 |
14.05 |
14.02 |
13.98 |
13.94 |
13.90 |
13.87 |
| 2 |
13.83 |
13.79 |
13.75 |
13.72 |
13.68 |
13.64 |
13.61 |
13.57 |
13.54 |
13.50 |
| 3 |
13.46 |
13.43 |
13.39 |
13.36 |
13.32 |
13.29 |
13.25 |
13.22 |
13.18 |
13.15 |
| 4 |
13.11 |
13.08 |
13.04 |
13.01 |
12.98 |
12.94 |
12.91 |
12.88 |
12.84 |
12.81 |
| 5 |
12.78 |
12.74 |
12.71 |
12.68 |
12.64 |
12.61 |
12.58 |
12.55 |
12.52 |
12.48 |
| 6 |
12.45 |
12.45 |
12.39 |
12.36 |
12.33 |
12.29 |
12.26 |
12.23 |
12.20 |
12.17 |
| 7 |
12.14 |
12.11 |
12.08 |
12.05 |
12.02 |
11.99 |
11.96 |
11.93 |
11.90 |
11.87 |
| 8 |
11.84 |
11.81 |
11.78 |
11.76 |
11.73 |
11.70 |
11.67 |
11.64 |
11.61 |
11.58 |
| 9 |
11.56 |
11.53 |
11.50 |
11.47 |
11.44 |
11.42 |
11.39 |
11.36 |
11.34 |
11.31 |
| 10 |
11.28 |
11.25 |
11.23 |
11.20 |
11.17 |
11.15 |
11.12 |
11.10 |
11.07 |
11.04 |
| 11 |
11.02 |
10.99 |
10.97 |
10.94 |
10.91 |
10.89 |
10.86 |
10.84 |
10.81 |
10.79 |
| 12 |
10.76 |
10.74 |
10.72 |
10.69 |
10.67 |
10.64 |
10.62 |
10.59 |
10.57 |
10.55 |
| 13 |
10.54 |
10.50 |
10.47 |
10.45 |
10.43 |
10.40 |
10.38 |
10.36 |
10.34 |
10.31 |
| 14 |
10.29 |
10.27 |
10.24 |
10.22 |
10.20 |
10.18 |
10.15 |
10.13 |
10.11 |
10.09 |
| 15 |
10.07 |
10.04 |
10.02 |
10.00 |
9.98 |
9.96 |
9.94 |
9.92 |
9.89 |
9.87 |
| 16 |
9.85 |
9.83 |
9.81 |
9.79 |
9.77 |
9.75 |
9.73 |
9.71 |
9.69 |
9.67 |
| 17 |
9.65 |
9.63 |
9.61 |
9.59 |
9.57 |
9.55 |
9.53 |
9.51 |
9.49 |
9.47 |
| 18 |
9.45 |
9.43 |
9.41 |
9.39 |
9.37 |
9.36 |
9.34 |
9.32 |
9.30 |
9.28 |
| 19 |
9.26 |
9.24 |
9.23 |
9.21 |
9.19 |
9.17 |
9.15 |
9.13 |
9.12 |
9.10 |
| 20 |
9.08 |
9.06 |
9.05 |
9.03 |
9.01 |
8.99 |
8.89 |
8.96 |
8.94 |
8.92 |
| 21 |
8.91 |
8.89 |
8.87 |
8.86 |
8.84 |
8.82 |
8.81 |
8.79 |
8.77 |
8.76 |
| 22 |
8.74 |
8.72 |
8.71 |
8.69 |
8.67 |
8.66 |
8.64 |
8.63 |
8.61 |
8.59 |
| 23 |
8.58 |
8.56 |
8.55 |
8.53 |
8.51 |
8.50 |
8.48 |
8.47 |
8.45 |
8.44 |
| 24 |
8.42 |
8.41 |
8.39 |
8.38 |
8.36 |
8.35 |
8.33 |
8.32 |
8.30 |
8.29 |
| 25 |
8.27 |
8.26 |
8.24 |
8.23 |
8.21 |
8.20 |
8.18 |
8.17 |
8.16 |
8.14 |
| 26 |
8.13 |
8.11 |
8.10 |
8.08 |
8.07 |
8.06 |
8.04 |
8.03 |
8.01 |
8.00 |
| 27 |
7.99 |
7.97 |
7.96 |
7.94 |
7.93 |
7.92 |
7.90 |
7.89 |
7.88 |
7.86 |
| 28 |
7.85 |
7.84 |
7.82 |
7.81 |
7.80 |
7.78 |
7.77 |
7.76 |
7.74 |
7.73 |
| 29 |
7.72 |
7.70 |
7.69 |
7.68 |
7.66 |
7.65 |
7.64 |
7.63 |
7.61 |
7.60 |
| 30 |
7.59 |
7.57 |
7.56 |
7.55 |
7.54 |
7.52 |
7.51 |
7.50 |
7.49 |
7.47 |
|
|