European Commission project  for your
Life & the Environment click here

Oxygen Monitors & water quality monitoring Instrumentation

Environmental & sustainable water treatment to save the planet

factory.jpg (3825 bytes)

cool hit counter

| Contact us  | About us Search |  Shop ||

Web  sites               | DrydenAqua AFM  |  Ozone | Membrane filtration | Swimming Pools
Instrumentation       | Index for instruments | Oxygen probes | AquaTraul O2 monitors |  Refractometers | bacterial dip slides | Total Gas |
Water test kits         | index for water test kits | Swimming pools | aquaculture & aquaria | drinking water | water treatment |

Oxygen Probe Manual

Typical Specifications:  
  • No zero point error
  • Very accurate usually better than +/-  0.2mg/l
  • Calibrate to 100% in air
  • Self-temperature compensating from 5 to 40 deg C
  • Connect with good quality two core cable,  tested up to 2000m with ordinary cable
  • Very stable, usually 3 to 6 months between calibration checks
  • Very easy to maintain, can be serviced in the field
  • Strong Teflon membrane
  • Very heavy duty sub-sea urethane cable, into epoxy sealed internals
  • Diameter = 63 mm, length = 73 mm. Cable length = 5 m (standard)
  • Weight 1000 g incl. cable 5m
  • 6 to 12 mill volt per ppm (mg/l) ( depends on temp)
  • pressure to 10 atmospheres
  • Water flow should be at least 4 cm/sec
  • Response time, to change in DO, approx 15 to 30 seconds
  • Standard probe is a 2 wire,  with temp sensor there are 5 wires
  • Temperature sensor PT100 built in to probe is available as an option
Oxygen_p.jpg (15105 bytes)
Calibration  

Calibration of the oxygen probe is very easy, just follow the steps below;

Standard calibration

  1. Remove probe from water, dry and clean the probe and membrane with a soft clean cloth.

  2. Using a pin, clear the breather hole on top of the probe

  3. Suspended the probe in air above the surface of the water, try to keep the probe out of the wind and direct sun-light. The probe should be left for a period of at least 30 minutes,  or until you are sure that the temperature of the probe is the same as the air temperature.  The best time to perform this task is early in the morning or late evening.

  4. The probes will now be reading the equivalent of 100% saturation. Take an accurate reading of the air temperature by the probes. Use the table below to establish the concentration in mg/l for the probe, and use this figure with your oxygen monitoring equipment.

  5. Example. if the air temp is 20.5 deg C,  and the probe is at the same temperature as the air,  the probe should be reading 8.99mg/l. Set the display to give this reading

Quick calibration

Sometime you do not have time to wait 30 minutes or more to perform a calibration, we therefore have a quick calibration technique that often proves more accurate than the standard technique.

  1. Remove the probe (or probes from the water), do not dry the probes, suspended in air out of direct sunlight

  2. Clear the breather hole with a pin and wipe the membrane with a soft cloth

  3. Measure the temperature of the water  ( if more than one probe is being calibrated,  the water temperature should be know for each probe)

  4. Calibrate the probes using the water temperature. The probes must be calibrated within 5 minutes of removing from the water, otherwise the temperature of the probe may have changed too much from the temperature of the water

 

 
Probe maintenance  

 

The oxygen probe requires very little attention,  the degree of attention will depend upon the water type in which the probe is immersed. For most applications we recommend that the probe is removed from the water once a week and the membrane cleaned with a soft clean cloth. The breather hole on top of the probe should also be cleaned using a pin. This is all that you need do with the probe on a regular basis.

In the event that the membrane is damaged,  the readings will become very erratic. Under these conditions the membrane should be replaced.

Membrane Replacement

The following procedure describes membrane replacement.

  1. Remove the oxygen probe from the water and clean with a cloth or paper towel

  2. Unscrew the bottom end cap, please note that the electrolyte containing a white deposit of colloidal solution of zinc oxide may be captained in the cap.   Discard this solution.

  3. Using a coin, unscrew the threaded retaining ring in the membrane cap, remove and discard the membrane and small `o` ring located below the membrane

  4. Clean the inside of the membrane cap thoroughly with a damp cloth, and finish off with a clean dry cloth.

  5. Insert a new `o` ring,  and then the membrane on top of the `o` ring. Screw down the threaded retaining ring until you feel tension, then give is a further 1/4 turn. If  the membrane wrinkles,  you will need to replace the `o` ring and membrane and try again.

  6. Clean the inside of the probe top, you may clean the silver cathode with 1500 grade paper,  take care as damage to the silver cathode can affect probe readings.  You can clean the angular zinc anode with rough abrasive paper in order to remove any oxidation layer. Clean the probe in freshwater to remove any of the fines.

  7. Fill the membrane cap with electrolyte, and holding the probe vertically, screw the membrane cap on to the top of the probe, making sure that the large `o` ring is in place. Slowly screw up the cap,  the excess electrolyte will escape though the breather hole.  Take care that you do not screw the cap on too quickly since this will stretch the membrane. Screw up the cap until it seals on the large `o` ring a then give it a further 1/4 turn.

  8. Immerse the probe into the water, after 24 hours perform a calibration as per the instructions above.

Components supplied with the probe

The standard  probe is supplied with he following components;

  1. Oxygen probe fitted with 5 meters of heavy duty cable

  2. 50ml bottle of electrolyte

  3. 5 spare membranes and membrane o rings

Note.

The probe is supplied dry without any electrolyte. On receipt of the probe, remove the end cap, fill end cap with electrolyte and slowly screw the cap back up onto the probe.  Screw the end cap up slowly to allow the excess electrolyte to squirt out of the top breather hole. You should then leave the probe for 30 minutes suspended in air before you perform a calibration. Once the probe is calibrated immerse the probe into the water for a period of at least 24 hours and then perform another calibration. Thereafter a calibration will be required at approximately 1 to 2 week intervals. After a period of approx. 4 weeks,  the calibration frequency can be reduced to once every month to several months.

Low Temperature applications

If the oxygen probe is subjected to sub-zero temperatures,  the electrolyte may freeze and cause damage to the probe.  In order to avoid freezing down to temperatures of -25 deg C glycerine is added to the electrolyte at a level of 55% glycerine to 45% water by volume. The glycerine will reduce the milli volt out-put of the probe down to approximately 30mv.

 

OxyGen Probe wire colours

 

Integration with your own system

The probe can be considered a low impedance millivolt generator. It has built in temperature compensation for readings only in mg/l. The output is linearly proportional to the oxygen concentration, and it is about 120 mv in air at a temperature of 20 deg C. The out-put is approx 10mv per mg/l of oxygen.

The output impedance is very low (< 1 kohm) so noise problems normally do not exist. However it is important that the input impedance of the transmitter is at least 3 milli ohms in order not to disturb the temperature compensation. It is also extremely important that inputs from probes are galvanically isolated from each other if more than one probe is connected to the same electronics.

The oxygen probe is easy to install. It should be placed where there is some movement in the water 2 cm/sec. Ensure that the probe cannot strike against the tank wall or similar) and don't mount it directly above air or oxygen diffusers etc.

Cable connections

Each probe is connected using ordinary 2-core cable and it is fitted with 5 meters of urethane cable unless otherwise specified. The brown ( or red)  wire is positive the blue (or white)  negative Use the junction box supplied with the probe when extending the cable.

With regards to the temperature sensor ( type pt100) the cable is numbered as follows;

Standard Oxygen sensor

+ve oxygen wire     (brown)

-ve oxygen wire      (blue)

Oxygen probe with pt100 sensor

1            +ve  dissolved oxygen ( red)

2            -ve  dissolved oxygen connection ( blue)

3            connected to the other side of the PT100 ( black)

4 & 5     joined together on one side of the PT100 ( green & yellow)

Suppression Circuit

The suppression circuit is used to drop the out-put of our oxygen probes from approximately 80mv to around 25mv.  The suppression circuit is required for oxygen meters such as Eutech units.

surpressing.BMP (222262 bytes)

 

Problem solving

 

In the unlikely event that you experience a problem with our oxygen probes,  then there are some simple checks that you can perform to identify and or solve the problem.

 

 

Problem Description Action
Zero or low reading A zero reading means that there is no voltage being generated by the probe, or there is a break in the electrical circuit. The milli volt out-put from the probe should be between 60 and 120mv in most cases, with the probe in air at 20 deg C  The out-put is approx 10mv per mg of oxygen. If the reading is very low check the following..  
Check that there is electrolyte in the probe change the membrane and electrolyte
Check that there is no cable damage If there is cable damage then cut out or repair the damage. Check that there is no water or moisture in junction boxes or cable connections.
biofouling of membrane can occur in in biologically active water. clean the membrane with a soft tissue once a week, or more frequently if required
High reading The milli volt out-put from the probe should be between 60 and 120mv in most cases, with the probe in air at 20 deg C  The out-put is approx 10mv per mg of oxygen.  If the reading is higher, check the following;

There should only be one small `o` ring below the membrane, if there are two `o` rings, or if the `o` ring is above the membrane then this will give the wrong tension on the membrane.

 

change the membrane  and electrolyte
The large `o` ring between the top and bottom sections of the the probe must be fitted,  if it is missing then there will be too high a tension on the membrane. change the membrane and electrolyte and fit a replacement large `o` ring
Check the silver cathode for any damage.  the cathode  should be clean and smooth without any indentations.  If the cathode has been damage, you will experience high erratic readings If there is only minor damage then using wet/dry paper, 700 grade gently remove the indentations.
If none of the above apply, remove the electrolyte from the probe, wash it electrode in fresh water an then thoroughly dry.  The probe should now give a zero reading  which will be less than 2 mv or 0.2mg/l of oxygen.  If the probe gives a reading above 2 or 3mv,  then it means that there is  a short out in the probe due to water ingress. If there is no damage to the probe cable, then the failure is internal. Return the probe to Dryden Aqua.
   
un-stable reading New probes It takes approximately 6 weeks for a new probe to fully condition stabilise.
Blocked breather hole. All probes have a small breather hole which should be kept clean. If the hole becomes blocked, gas inside the probe can inflate the membrane and push it off the cathode.  This can give high and or unstable readings clean the hole with a pin once a week.
If the prove is hitting the side of a tank,  the readings can become erratic position the probe such that it does not swing and hit the side of the tank.
Air or oxygen bubbles sitting below the membrane When the oxygen probe is used in aeration or oxygenation tanks, bubbles can sit on the underside of the membrane,  this can lead to false readings. Position the probe in a strong current of water, or tilt the probe by 45 deg.

 

 

Solubility of Oxygen in water

The solubility of oxygen in water in equilibrium with air at 760mm Hg pressure and 100% relative humidity Units:mg/1 . The data gives the mg/l of oxygen in solution that is equivalent to 100% saturation at the specified temperature

T deg C 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
0 14.60 14.65 14.52 14.48 14.44 14.40 14.36 14.33 14.29 14.25
1 14.21 14.17 14.13 14.09 14.05 14.02 13.98 13.94 13.90 13.87
2 13.83 13.79 13.75 13.72 13.68 13.64 13.61 13.57 13.54 13.50
3 13.46 13.43 13.39 13.36 13.32 13.29 13.25 13.22 13.18 13.15
4 13.11 13.08 13.04 13.01 12.98 12.94 12.91 12.88 12.84 12.81
5 12.78 12.74 12.71 12.68 12.64 12.61 12.58 12.55 12.52 12.48
6 12.45 12.45 12.39 12.36 12.33 12.29 12.26 12.23 12.20 12.17
7 12.14 12.11 12.08 12.05 12.02 11.99 11.96 11.93 11.90 11.87
8 11.84 11.81 11.78 11.76 11.73 11.70 11.67 11.64 11.61 11.58
9 11.56 11.53 11.50 11.47 11.44 11.42 11.39 11.36 11.34 11.31
10 11.28 11.25 11.23 11.20 11.17 11.15 11.12 11.10 11.07 11.04
11 11.02 10.99 10.97 10.94 10.91 10.89 10.86 10.84 10.81 10.79
12 10.76 10.74 10.72 10.69 10.67 10.64 10.62 10.59 10.57 10.55
13 10.54 10.50 10.47 10.45 10.43 10.40 10.38 10.36 10.34 10.31
14 10.29 10.27 10.24 10.22 10.20 10.18 10.15 10.13 10.11 10.09
15 10.07 10.04 10.02 10.00 9.98 9.96 9.94 9.92 9.89 9.87
16 9.85 9.83 9.81 9.79 9.77 9.75 9.73 9.71 9.69 9.67
17 9.65 9.63 9.61 9.59 9.57 9.55 9.53 9.51 9.49 9.47
18 9.45 9.43 9.41 9.39 9.37 9.36 9.34 9.32 9.30 9.28
19 9.26 9.24 9.23 9.21 9.19 9.17 9.15 9.13 9.12 9.10
20 9.08 9.06 9.05 9.03 9.01 8.99 8.89 8.96 8.94 8.92
21 8.91 8.89 8.87 8.86 8.84 8.82 8.81 8.79 8.77 8.76
22 8.74 8.72 8.71 8.69 8.67 8.66 8.64 8.63 8.61 8.59
23 8.58 8.56 8.55 8.53 8.51 8.50 8.48 8.47 8.45 8.44
24 8.42 8.41 8.39 8.38 8.36 8.35 8.33 8.32 8.30 8.29
25 8.27 8.26 8.24 8.23 8.21 8.20 8.18 8.17 8.16 8.14
26 8.13 8.11 8.10 8.08 8.07 8.06 8.04 8.03 8.01 8.00
27 7.99 7.97 7.96 7.94 7.93 7.92 7.90 7.89 7.88 7.86
28 7.85 7.84 7.82 7.81 7.80 7.78 7.77 7.76 7.74 7.73
29 7.72 7.70 7.69 7.68 7.66 7.65 7.64 7.63 7.61 7.60
30 7.59 7.57 7.56 7.55 7.54 7.52 7.51 7.50 7.49 7.47

 

 

CORRECTION FACTORS FOR CALIBRATION OF A DISSOLVED OXYGEN

 

In order to calibrated the oxygen probe for use in seawater or higher salinity application, you need to make a compensation for the salinity of the water.  Calibrate the probe as per normal according to the solubility tables  ( example, 20.5 deg C =8.99mg/l = 100% sat) . If the salinity is 30ppt, the correction factor is 83%, the reading on the oxygen meter should then be 8.99 x 0.83 = 7.46mg/l = 100% sat in seawater at 20.5 deg C and salinity of 30ppt.
Salinity ppt  of the water in which the oxygen meter is to be used Dissolved Oxygen setting, correction factor
0 100%
5 97%
10 94%
15 91%
20 88%
25 85%
30 83%
35 80%

VARIATION OF MEAN ATMOSPHERIC PRESSURE WITH ALTITUDE

 

Once the probe has been calibrated at salinity correction factors applied,  you may also set a correction factor for the barometric pressure.  This aspect of probe calibration is usually ignored.  However if you are located at high altitude then it will be an important consideration.
(m)height Average Atmospheric Pressure (mmHg) Correction Factor
0

760

1.00
100

750

1.01
200

741

1.03
300

732

1.04
400

723

1.05
500

714

1.06

 

Copyright © 1997
Last modified: March 26, 2008